LIGHT PROGRESS
FLOW-C-NX [Eng.]
Rev. 2 Oct -2020
Pag. 11/16
MAINTENANCE
UV LAMP
USE
PROTECTIVE
GLOVES
WEAR
PROTECTIVE
GLASSES
Lamp maintenance is very important because UV-C light emitted by the source is not visible to the eye, so,
although you see the lamp ON, its germicidal effect may not be sufficient to performs disinfection treatment.
Before reaching the lamp,
ALWAYS
switch off power supply
If UV-C lamp gets dirty (or gets dusty) clean it with a clean cloth and alcohol, and avoid touching the glass part with
hands. To check UVC lamp life number of hours, correct product code of the spare UV lamp please refer to table 1.
The lamp, once it has reached the timing for replacement, must
be replaced
, this operation is carried out as follows:
•
For each side of the device, unscrew one screw (or knob) which sets the screen and rotate it 90 degrees.
•
Loosen the nuts of the holder and slide them along the lamp, grab the lamp from the 2 ends and pull it in a di-
rection parallel to the unit.
•
Repeat steps in reverse to reassemble the new lamp again.
TROUBLESHOOTING
UV LAMPS
See MIANTENANCE
BALLAST
All UV models are supplied with electronic ballasts which guarantee an emission of 15% more if compared with traditional
ballasts. Furthermore, lamps are not influenced by environment temperature, (<-20°C). Finally, the switching on is
immediate, with no starter, and for this reason lamps last longer.
To preserve the original characteristics of IP sealing and insulation, in order to keep intact all the characteristics necessary
to ensure the long-lasting life of the product, the replacement of the electronic ballast must be carried out only by trained
personnel with requirements recognized by LIGHT PROGRESS; if necessary, you should return the product to the
company for repair.
Alternatively, you can purchase the entire power supply duct for substitution (for code see Table n.1).