Lighthouse ApexRemote Operating Manual
6-2
248083447-1 Rev 1
Running the
Instrument
Using
MODBUS
The applicable action commands are displayed in Table 6-1.
Each of the described action commands above are written to the
command register (40002).
AUTOMATIC Counting Mode
In Automatic counting mode, the instrument uses the configured
sample time and hold time to record samples.
The instrument will continue running samples at the configured sample
time until it receives a stop command. When the stop command is
given, any partial data will not record to the buffer.
After setting all the instrument parameters as described in “Changing
the Default Instrument Parameters” on page 6-4, write these commands
to the Command register (40002):
11
Start Instrument; to start recording
12
Stop Instrument; to stop recording
Table 6-1 Action Commands
Value
Action
1
Saves all writable 4xxxx register values to the EEPROM.
3
Clears the Data Buffer. Record count is set to zero.
4
Saves the instrument parameters in the 40xxx registers to
the EEPROM. Parameters include Sample Time, Hold
Time, and Location.
11
Instrument Start (Automatic Counting).
Uses defined Hold Time and Sample Time. Instrument
executes samples and holds until an Instrument Stop
command is issued.
12
Instrument Stop.
Aborts current sample. Stops data collection.
Summary of Contents for ApexR02
Page 1: ......
Page 2: ...00 ...
Page 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Page 6: ......
Page 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Page 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Page 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Page 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Page 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Page 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Page 83: ...00 ...
Page 84: ......