Lighthouse ApexRemote Operating Manual
6-8
248083447-1 Rev 1
register 30016’s threshold register would be 45016.
Setting the Alarm Threshold Value
The Alarm Threshold Value is set in the low register of the channels.
Each channel has independent threshold value registers. Since any or
all channels can be enabled for alarms at any given time, each threshold
value applies to the corresponding channel. Setting a value for channel
1 as 100 will not affect channel 2 setting of, say, 500. See Table 6-9.
Table 6-8 Alarm Threshold Registers
Register Data
Type
Description
45009
unsigned int Threshold for Particle Channel 1 [high]
(smallest particle size starts here)
45010
unsigned int Threshold for Particle Channel 1 [low]
45011
unsigned int Threshold for Particle Channel 2 [high]
45012
unsigned int Threshold for Particle Channel 2 [low]
45013
unsigned int Threshold for Particle Channel 3 [high]
45014
unsigned int Threshold for Particle Channel 3 [low]
45015
unsigned int Threshold for Particle Channel 4 [high]
45016
unsigned int Threshold for Particle Channel 4 [low]
Table 6-9 Alarm Threshold Registers set to 1000
Registers
Particle
Channel
Threshold
Value
45009 - 45010
1
1000
45011 - 45012
2
1000
Summary of Contents for ApexR02
Page 1: ......
Page 2: ...00 ...
Page 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Page 6: ......
Page 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Page 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Page 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Page 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Page 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Page 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Page 83: ...00 ...
Page 84: ......