Operation - Mass Concentration Mode
5HY
5-43
PRINT RANGE (BUFFER REPORT)
Note:
PARTICLE data
may be printed, even if the
data was taken in MASS
mode. On the MAIN
Screen, press the MASS
button, which changes it to
PARTICLE. Return to this
point to print the particle
counts.
The instrument stores up to 3000 records in its data buffer. The Buffer
Report prints all or a range of the records in the buffer in the format
specified by the selections in the Configuration.
At the end of the Buffer Report, the following summary information is
printed.
•
Data is displayed in PM or UGR mode
•
# of samples in the report
•
For each channel size:
•
Maximum Value
•
Minimum Value
•
Average
•
Standard Deviation
The UCL calculation will be printed for each channel size if the
number of locations in the range is 2 through 9. The calculation will
only print if the Format settings (Setting screen) are set to CUML and
NORM. Any other format settings or locations less than 2 and greater
than 9 will print N/A for each channel.
View the same data in different formats just by changing the Particle
Display options on the Sample Setting screen.
The Data Summary can be viewed as:
•
PM
•
UGR/m3
Printing Buffer Report
•
To print a range of data from the buffer, scroll to the first data
record to be included in the range.
•
Press the RANGE button. The following screen will display as
shown in Figure 5-50.
Summary of Contents for Handheld 3016 IAQ
Page 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Page 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Page 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Page 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Page 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Page 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Page 135: ...00...