Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual
5-2
5HY
Using the
Instrument
The HANDHELD comes with a charged battery and is ready for use.
To start using the instrument, proceed as follows:
WARNING:
Do not
attempt to sample
reactive gases (such as
hydrogen or oxygen)
with this instrument.
Reactive gases create an
explosion hazard in the
instrument.
Sampling any gas under
pressure can damage the
instrument and void the
warranty.
Sampling any gas that is
not the same density as
ambient air can result in
inaccurate data.
Contact Lighthouse for
more information.
1.
Insert the battery included with the shipment (see Chapter 2 for
instructions.)
2.
Position the instrument in the environment to be measured.
3.
Remove the protective cap from the inlet tube. To use the
provided isokinetic probe, install it by connecting to inlet tube on
the top of the instrument.
NOTE: Do not discard the protective
cap. It should be placed on the inlet tube any time the
instrument is to be moved outside the environment being
measured.
4.
The included Temperature/Relative Humidity probe can be
attached to the provided receptacle to read environmental data.
5.
Set on/off switch found on the left side of the unit to ON.
6.
The Start Up screen displays on the LCD.
7.
The MAIN screen appears.
WARNING:
Do NOT
allow water, solvents, or
other liquids enter the
instrument via the inlet
tube - the instrument will
be damaged and void the
warranty.
Do NOT operate the
instrument with the inlet
tube capped or plugged -
the internal pump will be
damaged and void the
warranty.
8.
On the touch screen, press the START button to start the
instrument.
9.
“STARTING” will display when the pump is initially turned on.
10. When the HANDHELD starts counting, “COUNTING” appears
on the display. Data is displayed according to the size of each
particle.
11. If the instrument is in AUTO mode with cycles and a hold time,
“HOLDING” will display after each cycle and “FINISHED” will
display when all the cycles are complete.
12. Press the “STOP” button to stop the instrument before the cycles
are complete.
Note:
The pump motor
requires a minimum 5
seconds to ramp up to full
flow. Refer to “Pump
Startup” on page 5-26.
Summary of Contents for Handheld 3016 IAQ
Page 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Page 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Page 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Page 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Page 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Page 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Page 135: ...00...