Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual
5-16
5HY
SETTINGS
The instrument can be configured to count in different modes and
formats. See Figure 5-15.
Figure 5-15 Sample settings screen
COUNT MODE
The following modes are available: Auto and Manual.
•
AUTO
- When the instrument is in Automatic Mode and the
START button is pressed, the instrument will start counting
particles automatically according to the Sample Time, Hold Time
and the number of Cycles that are configured. If Cycles are set to 0,
the instrument will continue indefinitely in Auto Mode until the
STOP button is pressed.
•
MANU (Manual Mode)
- When the instrument is in Manual
Mode, it will start counting when START is pressed and stop at the
end of one programmed Sample Time.
PARTICLE Display.
The data format is either PM or Mass Concentrations (UGR). PM
indicates Particulate Matter, expressed in micrograms per cubic meter
(ugr/m3). PM is calculated by summing the mass concentration levels
for all channels smaller than the indicated channel. For example, the
PM 1.0 = (Mass Concentration of 0.3) + (Mass Concentration of 0.5).
PM 10.0 = the sum of the Mass Concentration levels of the 0.3 + 0.5 +
1.0 + 2.5 + 5.0
m channels. Note that the Mass Concentration of the
10.0
m channel is not included. It is, however, included in the TPM
value, the Total Particulate Matter contained in all channels.
Summary of Contents for Handheld 3016 IAQ
Page 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Page 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Page 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Page 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Page 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Page 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Page 135: ...00...