Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
5-2
248083408-1 Rev 1
5.
Configure the unit via the user interface to sample for 5 minutes
and set a 10 second hold.
6.
Set Cycles to 10 so the instrument will take 10 five-minute
samples.
7.
If an average of more than one count per five minute sample is
reported, reset the instrument to sample for 30 minutes again to
purge it, then repeat the Purge Test again.
8.
After the instrument has met the requirement of the Purge test,
return the instrument to its normal location and operating status.
If the instrument still fails the Purge Test, contact Lighthouse Tech
Support for assistance.