Lighthouse REMOTE 2014P, 3014P, 5014P Operating Manual
A-16
248083408-1 Rev 1
Data Type Registers
Note:
All data records
have the same data types
assigned to them. The
user does not have to read
the data type registers for
every record.
The 41xxx register series is used to identify the type of data items in the
30xxx series. The Data Type registers run in parallel with the Data
Registers. For example, Data Register 30041’s Data Type register is
41041.
Data Types are assigned 4 ASCII characters across 2 registers. If a
Data Type string contains less than 4 characters, then the rest of the
string is padded with NULL characters. Note that a Data Type using all
four characters will not end with a NULL character.
Table A-15 Data Types
String
Description
TIME
Timestamp
STIM
Sample Time
SVOL
Sample Volume
LOC
Location
STAT
Status
TEMP
Temperature
RH
Relative Humidity
AIRV
Air Velocity
DPRS
Differential Pressure
ESD
Electrostatic Discharge
FLOW
Flow Rate
LASV
Laser Voltage
VOLT
Voltage
PRES
Pressure
Summary of Contents for Remote 2014P
Page 1: ......
Page 2: ......
Page 6: ......
Page 10: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual t iv 248083408 1 Rev 1...
Page 12: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual ii 248083408 1 Rev 1...
Page 14: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 1 2 248083408 1 Rev 1...
Page 20: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 2 6 248083408 1 Rev 1...
Page 40: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 3 20 248083408 1 Rev 1...
Page 72: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual B 2 248083408 1 Rev 1...
Page 74: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual C 2 248083408 1 Rev 1...
Page 79: ......
Page 80: ......