Lighthouse REMOTE-2 PN Series Operating Manual
6-2
248083310-1 Rev 3
4.
Allow the instrument to sample through a 30 minute period. This
time allows the unit to warm up and purge any residual particles
that might be inside it.
5.
Configure the unit via the user interface to sample for 5 minutes
and set a 10 second hold.
6.
Set Cycles to 10 so the instrument will take 10 five-minute
samples.
7.
If an average of more than one count per five minute sample is
reported, reset the instrument to sample for 30 minutes again to
purge it, then repeat the Purge Test again.
8.
After the instrument has met the requirement of the Purge test,
return the instrument to its normal location and operating status.
9.
If the instrument still fails the Purge Test, contact Lighthouse
Tech Support for assistance.
Summary of Contents for REMOTE 3012PN
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 2 PN Series Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 2 PN Series Operating Manual t iv 248083310 1 Rev 3...
Page 12: ...Lighthouse REMOTE 2 PN Series Operating Manual ii 248083310 1 Rev 3...
Page 14: ...Lighthouse REMOTE 2 PN Series Operating Manual 1 2 248083310 1 Rev 3...
Page 42: ...Lighthouse REMOTE 2 PN Series Operating Manual 4 14 248083310 1 Rev 3...
Page 46: ...Lighthouse REMOTE 2 PN Series Operating Manual 5 4 248083310 1 Rev 3...
Page 50: ...Lighthouse REMOTE 2 PN Series Operating Manual A 2 248083310 1 Rev 3...
Page 52: ...Lighthouse REMOTE 2 PN Series Operating Manual LOF 2 248083310 1 Rev 3...
Page 56: ...Lighthouse REMOTE 2 PN Series Operating Manual I 4 248083310 1 Rev 3...
Page 57: ...00...
Page 58: ......