Lighthouse REMOTE Active Count Installation Guide
5-2
248083401-1 Rev 5
Running the
RAC
The Remote ActiveCount can be run by MODBUS Protocol. The
applicable action commands are displayed in Table 5-2.
Access RAC
Information
Table 5-2 Action Commands
Value
Register
Action
1
40002
Saves all writable 40002 register values to the
EEPROM.
5
40002
Enable Remote Mode. Locks out the instrument’s user
interface. Can only change instrument parameters via
MODBUS.
11
40002
Instrument Start. Uses defined Initial Delay, Hold
Time, Sample Interval and counting mode. Instrument
executes samples and holds until an Instrument Stop
command is issued. This command will start the
pump.
12
40002
Instrument Stop. Aborts current sample. Stops pump.
Stops data collection.
Table 5-3 Read Commands
Value
Register
Action
50
40001
Read Device Information: Parse through device
information. Verify “VIABLE” in the product name.
1
40003
Read Device Status
2
40005
Read Device Serial Number
74
30001
Get Data Record
72
41001
Get Data Types
72
42001
Get Data Units
72
43001
Get Data Enables
Summary of Contents for REMOTE Active Count
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Page 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Page 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Page 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Page 61: ...00...
Page 62: ......