Get Started
248083227-1 Rev 8
3-7
1. Make sure that there is no flow of liquid in the system and power is
removed from the RLPC.
Note:
When connecting
flare fittings to the RLPC,
do
NOT
use a wrench or
other tool to tighten the
flare nut! Doing so will
damage the instrument,
tubing or nut. The
connections
MUST
be
finger-tight
only.
2. Connect the liquid source to the RLPC’s Inlet nozzle using the
provided ¼" flare tubing. Make sure the tubing is inserted firmly
onto the nozzle and the knurled nut is finger tight only.
3. Connect the liquid drain to the RLPC’s Outlet nozzle using the
provided ¼" flare tubing. Make sure the tubing is inserted securely
onto the nozzle and the knurled nut is finger tight only.
4. Turn the liquid system on and allow liquid to flow through the
RLPC, making sure that there are no leaks at the sensor’s inlet or
outlet or any other connections.
WARNING:
If a leak
occurs anywhere within
the system, immediately
stop the flow of liquid
and repair the leak!
5. If leaks occur, immediately stop the flow of liquid. Disconnect and
clear the flare fittings and connectors of debris or burrs. Repair any
other leak(s) and return to Step 1 of this section.
6. If no leaks occur, connect power to the RLPC.
7. Make sure the sample liquid is bubble-free - bubbles will be
counted as particles and in large amounts may cause sensor errors.
8. Allow liquid to flow through the sensor for 10 minutes or until
there are no bubbles and the Service LED is OFF. If the Service
LED comes on and remains on, the flow cell should be cleaned (see
Appendix C, “Maintenance”
for cleaning instructions).
9. If there are no leaks and the service LED is off, the instrument is
ready to configure the communications - refer to
Chapter 4,
“Communications”
.
10. Adding devices or additional connections between the sample
source and the sensor will increase the chances for leaks and
bubbles - keep connections to a minimum.
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......