Communications
248083227-1 Rev 8
4-3
DIP Switch Addressing
Note:
Because Address
0 is reserved for
broadcasting in MODBUS
RS485 communications,
Address 1 is set both when
all the DIP switches are
OFF or when DIP switch1
is ON.
For MR Protocol, Address
0 is set when all the DIP
switches are OFF. Use a
thin, pointed tool such as a
small jeweler’s screwdriver
or a straightened paper clip
to set the switches.
Table 4-3 shows the addresses set by DIP switches 1-6.
Table 4-3 DIP Switch Addressing
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
0 0 0 0 0 0
0 or 1
0 0 0 0 0 1
32
1 0 0 0 0 0
1
1 0 0 0 0 1
33
0 1 0 0 0 0
2
0 1 0 0 0 1
34
1 1 0 0 0 0
3
1 1 0 0 0 1
35
0 0 1 0 0 0
4
0 0 1 0 0 1
36
1 0 0 1 0 0
5
1 0 1 0 0 1
37
0 1 0 1 0 0
6
0 1 1 0 0 1
38
1 1 1 0 0 0
7
1 1 1 0 0 1
39
0 0 0 1 0 0
8
0 0 0 1 0 1
40
1 0 0 1 0 0
9
1 0 0 1 0 1
41
0 1 0 1 0 0
10
0 1 0 1 0 1
42
1 1 0 1 0 0
11
1 1 0 1 0 1
43
0 0 1 1 0 0
12
0 0 1 1 0 1
44
1 0 1 1 0 0
13
1 0 1 1 0 1
45
0 1 1 1 0 0
14
0 1 1 1 0 1
46
1 1 1 1 0 0
15
1 1 1 1 0 1
47
0 0 0 0 1 0
16
0 0 0 0 1 1
48
1 0 0 0 1 0
17
1 0 0 0 1 1
49
0 1 0 0 1 0
18
0 1 0 0 1 1
50
1 1 0 0 1 0
19
1 1 0 0 1 1
51
0 0 1 0 1 0
20
0 0 1 0 1 1
52
1 0 1 0 1 0
21
1 0 1 0 1 1
53
0 1 1 0 1 0
22
0 1 1 0 1 1
54
1 1 1 0 1 0
23
1 1 1 0 1 1
55
Summary of Contents for RLPC 0.1m
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Page 6: ......
Page 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Page 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Page 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Page 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Page 117: ...00...
Page 118: ......