Configuring with LWS ASCII Service Protocol
248083227-1 Rev 8
5-9
Example: Setting Sample Time <ta#>
WARNING:
Do NOT
set the Sample Time to a
value less than 30
seconds.
The REMOTE instrument was preconfigured at the factory for 1
minute sample time. Do not set the value to less than 30 seconds. To
view the current sample time on the instrument in ASCII mode,
1.
Type:
<?>
This command displays the current sample time among
the other parameters.
To change the sample time to 2 minutes (120 seconds):
2.
Type:
<ta120>
Response:
<ta120> 00:02:00
To change the sample time to 1 minute (60 seconds):
3.
Type:
<ta60>
Response:
<ta60> 00:01:00