Programming
248083296-1 Rev 3
4-9
Command Set
VERBOSE MODE
Note:
Upper case V
<V>
Normally, VERBOSE MODE is off and the computer responds with
<OK>
after each command is received and correctly implemented.
Turning on VERBOSE MODE tells the computer to echo back the
information to the screen.
VERBOSE MODE is required in order to program the device. It is not
required to run the counter.
GET CURRENT SETTINGS
<?>
Shows current parameter settings.
Type:
<?>
Response:
<?>
Flow Rate = 0.1 cfm
Sample Time = 5 secs
Range1= 10000.
Range2= 1000.
Service High = 3000 mV
Service Low = 145 mV
Alarm Channel = 1
Alarm Threshold = 1000 counts
Suppress Alarms = 2
Channel 1 = 0.3
Channel 2 = 0.5
DISPLAY MENU OF COMMANDS
<m1>
Shows menu of commands.