Operating Counter
248083389-1 Rev
6-75
Printing a Report
To print a report, the user can choose to either clear the data before
starting the test or run the samples and then select the range of data to
include in the report.
•
Once the user has collected the necessary samples, press REPORTS
from the MAIN screen.
•
Press the RANGE button on the REPORTS screen
•
Using the ARROW buttons, scroll to the first record in the range of
data for the report.
•
Press the RANGE button on the Data View Screen.
•
Type in the number of records to be included in the range.
•
Press ENTER.
•
The display will return to the main REPORTS screen and start
printing the report.
Note:
If CANCEL is
pressed, the report may not
stop printing right away. It will
continue until there is a break
in the report processing.
•
To stop printing, press the CANCEL button.
Report Requirements
Fed Std ft
3
This report requires a minimum of 2 locations be measured, at least 1
sample be taken per location and at least 5 samples total be taken in
each cleanroom at a minimum sample volume of 0.1 cubic feet.
The 95% Upper Confidence Limit (UCL) will be calculated by the
SOLAIR if the number of sampled locations is equal to or less than 9.
Air flow (Unidirectional vs. Non-unidirectional) will have an effect on
the minimum locations and minimum samples required. In general, the
Non-unidirectional option requires more of both locations and samples.
For the Fed Std ft
3
report, the classification for a particle size is based
on its samples taken from all locations in the cleanroom. Each particle
size is given one of the following results:
•
a classification rating which will be one of the possible
classifications (1, 10, 100, 1000, 10000, 100000)
•
“Unable to Classify” if it fails one of the above requirements
•
an “Out of Range” if the calculated classification at that particle
count exceeds 100,000
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......