MODBUS Register v1.48
248083389-1 Rev
B-9
Data
Registers
Data is stored in the input registers (30xxx series), which are read-only.
All data items are four bytes long and are stored across two registers.
Byte and word order for integer data is big-endian. Thus, data items
are formed by placing the high bytes in front of the low bytes.
Example:
<High Bytes><Low Bytes> = <4 Byte Data Item>
IEEE floating point has big-endian byte order and little-endian word
order. Thus, analog data items are formed by placing the low bytes in
front of the high bytes.
Example:
<Low Bytes><High Bytes> = <4 Byte Data Item>
Not all particle and analog channels are necessarily active. Retrieving
data from an inactive channel returns garbage. See the Data Enable
Registers section of this document for details on how to record data
from active channels.
This entire series of registers represents one data record in the device.
The Record Index Register (40025) must be changed to index other
records here.
The first record in the data buffer is located at Index=0. The most
recently saved value is at Index=-1.
Table B-8 Data Registers
Register
Data Type
Description
30001
signed integer
Timestamp [high] (# of seconds
since midnight, 1/1/1970)
30002
signed integer
Timestamp [low]
30003
unsigned integer
Sample Time [high] (In seconds)
30004
unsigned integer
Sample Time [low]
30005
signed integer
Location [high] (Place where data
was recorded)
30006
signed integer
Location [low]
30007
unsigned integer
Data Status [high]
30008
unsigned integer
Data Status [low]
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......