Lighthouse S3100/S5100/Boulder Counter Operating Manual
B-12
248083389-1 Rev
Alarm Flags in Channels (30076)
Register 30076 represents the flag bits corresponding to particle
channels that have exceeded the threshold [Threshold High Registers
(45xxx series)] based on alarm mode
Data Type Registers
Note:
All data records
have the same data types
assigned to them. The
user does not have to read
the data type registers for
every record.
The 41xxx register series is used to identify the type of data items in the
30xxx series. The Data Type registers run in parallel with the Data
Registers. For example, Data Register 30041’s Data Type register is
41041.
Data Types are assigned 4 ASCII characters across 2 registers. If a
Data Type string contains less than 4 characters, then the rest of the
string is padded with NULL characters. Note that a Data Type using all
four characters will not end with a NULL character.
Table B-10 Data Types
String
Description
TIME
Timestamp
STIM
Sample Time
SVOL
Sample Volume
LOC
Location
STAT
Status
TEMP
Temperature
RH
Relative Humidity
AIRV
Air Velocity
DPRS
Differential Pressure
ESD
Electrostatic Discharge
FLOW
Flow Rate
LASV
Laser Voltage
VOLT
Voltage
PRES
Pressure
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...