Operating Counter
248083389-1 Rev
6-25
Concentration data will be recorded and can be viewed in the Data
screen and on the printouts.
Note:
BEEP mode only
works with CUMULATIVE
data and for Sample Times
equal to or greater than 6
seconds.
•
BEEP
- In this mode, the instrument is pre-configured to beep
according to the alarm threshold set in the Alarm configuration and
the instrument’s sample time when the instrument is set to collect
cumulative data. Counting starts one second after the START
button is pressed and will continue until the STOP button is pressed
or the number of configured cycles is reached.
If no channel is set for alarming, when BEEP mode is selected the
instrument will automatically select the smallest channel size, at
whatever the alarm threshold is set to, to alarm on.
If alarming is enabled on more than one channel, when the
instrument is in BEEP mode it will beep if the alarm threshold is
exceeded on any of the channels enabled for alarming.
Note:
If BEEP mode is
set to one count, each
beep may not be for every
single count.
The data will be recorded, based on the set sample time and hold
time, and can be viewed in the View Buffer and on the printouts,
however there will be no indication on the record that the data was
saved while the instrument was in BEEP mode.
Geiger Counter Mode
While in BEEP mode the instrument can be set to beep 1 to 4 beeps per
second depending on the number of particles the instrument has
counted per second.
To set up the instrument for Geiger Counter Mode:
1. Enable BEEP Mode.
2. Set the smallest particle channel to 1.
In Geiger Counter Mode the instrument will beep from 1 beep up to
4 beeps per second if it counts from 1 particle up to 4 particles per
second. The maximum beeps per second the instrument can emit
are 4 per second even if the instrument counts 100 particles per
second.
The value set in the smallest particle channel’s alarm is the trigger
for a single beep per second. If the smallest particle channel is set to
2, the instrument will beep once every second for every 2 particles
the instrument counts per second.
Summary of Contents for solair 3100
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Page 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Page 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Page 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Page 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Page 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Page 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Page 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Page 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Page 187: ...00...
Page 188: ......