Suzhou  Lily  Tech.  Co.,  Ltd. 

                                                                                                                                                                                                             

Page  1, Total  4  Pages

 

ZL-6231A Temperature Controller 

Instruction Manual   Version 1.2 

 

Introduction 

ZL-6231A thermostat is for cooling or heating control, with IP65 front panel, buzzer beeping and warning function. 

There is a multifunction timer which is for incubator air exhaustion, or egg tray turning, or other application. 

 

Main Specification 

Power supply: 185~245Vac, 50/60Hz 

Sensor: NTC, wire length is 2 meters 

Output: 10A, 250Vac 

resistance

 

Setting range: -40~120

 

Display range: -40~130

 

Working: -10~45

, 5~90%RH without dew   

Dimension: W78 x H34.5 x D71 (mm)

 

 

Installation drilling: W71 x H29 (mm) 

Case materials: PC + ABS fire proof 

Protection level: IP65 (Front side only) 

 

Keypad and Display 

Icon 

Function 

On 

Off 

Blinking 

 

R1 output 

R1 Temp. Output energized 

R1 Temp. Output de-energized 

Within protecting delay 

 

Work mode 

Cooling mode 

Not in cooling mode 

Setting set-point 

 

Work mode 

Heating mode 

Not in heating mode 

Setting set-point 

 

Maintenance 

---- 

No failure 

Has failure 

 

Warning 

---- 

No warning 

Has warning 

 

R2 output 

R2 energized 

R2 de-energized 

 

 

Egg turn times reached 

 

 

After reached, stop egg turn 

 

Display 

Remark 

E01 

Sensor failure (short or open) 

Hi 

Room temp. is higher than the high limit 

Lo 

Room temp. is lower than the low limit 

EE 

Memory error 

Err 

Password error 

i

i

A

A

 

 

E

E

x

x

t

t

e

e

r

r

n

n

a

a

l

l

 

 

w

w

a

a

r

r

n

n

i

i

n

n

g

g

 

 

U

U

n

n

L

L

 

 

Parameters Will restore to factory default settings

 

 

 

Power up Display 

    The model (ZL-6231A) and software version (Version 1.1) will display successively: “6231”, “A 1.2”.   

 
Operation 

Set Set-Point (Factory Default Setting = 37.8

)

 

Keep

S

depressed for 3 seconds to enter into the status. The current set-point value displays. 

Press

or

  to set the value

 

(keeping depressed can fast set).   

Press

S

to exit, and save the settings.   

The status will exit, and the setting will not be saved, if no key operation within 30 seconds. 

 

Set System Parameters 

Keep

P

depressed for 3 seconds, digits show “---0”.   

Press

to select the digit of the password, press

to set the value of the digit. 

Press

S

  to confirm: 

If the password is correct, enter into the parameter setting status, else display “Err”, and exit. 

Set in parameter setting status: 

Summary of Contents for ZL-6231A

Page 1: ...lure Has failure Warning No warning Has warning R2 output R2 energized R2 de energized Egg turn times reached After reached stop egg turn Display Remark E01 Sensor failure short or open Hi Room temp is higher than the high limit Lo Room temp is lower than the low limit EE Memory error Err Password error i iA A E Ex xt te er rn na al l w wa ar rn ni in ng g U Un nL L Parameters Will restore to fact...

Page 2: ...it_protection 2 U45 Timer repeat times 0 9999 If 0 Timer never stops Else R2 never on when U46 reaches U45 0 U46 Timer counter the value saved to permanent memory once every hour keeps even without power supply 0 9999 One period timer period 1 timer period 2 Clear to 0 before every hatching 0 U52 Over temp warning delay 0 180 min 0 U53 1st over temp warning delay after power supplied 0 180 hour 0 ...

Page 3: ...ailure or external warning input If U62 0 no buzzing warning If U62 1 there will be continuous buzzing of warning The waring will stop if press P or warning condition disappears Over Temperature Warning When Troom SP U54 or Troom SP U55 display warning and buzzing External Warning Input NO normal open If open no warning if closed warning NC normal close If closed no warning if open warning Locked ...

Page 4: ...dule as load or tungsten lamp or motor following the below requirements to avoid surging current damaging or shorten the life time of the controller outputs For ac dc module as load the rated current should be no more 1 10th of output specification under pure resistance For tungsten lamp as load the rated current should be no more 1 15th of output specification under pure resistance For motor the ...

Reviews: