Page 17
Display Functions and Indications
expected. This can only happen if there is an internal malfunction. If
resetting does not restore normal operation, contact Limerick, as the
pump may need to be returned for repair.
Internal Fault:
This error code means an internal malfunction has occurred. Contact
Limerick as the pump may need to be returned for repair.
Vacuum Time Out:
This fault could be caused by the following:
•
Air leak in accessory kit
•
Mother was not setup to pump within appropriate time after
pump was started (must have breast cup attached within 7
seconds of starting pump).
•
When encountering this problem, manually restart the pump by
turning the pump off and back on using the Cycle/Power knob (1).
(
See the illustration on page 11
)
•
If the problem persists, contact Limerick as the pump may need to
be returned for repair.
Other problems that may occur
without a fault message
Situation “A”
If tubing becomes kinked or milk back flows into the filter the pump
will lose vacuum and pulsate quickly. You will either need to unkink
tubing or change the filter if milk is in the filter, as the filter cannot
be cleaned. (Filter is a safety feature that prevents milk, bacteria, and
viruses from getting into the motor of the pump.) It’s important to
Summary of Contents for Joy
Page 1: ...Owners Manual ...