11
1.
Insert the multi socket into the mA and HzV
Ω
terminals.
2.
Set the rotary switch to an appropriate measurement position in F range.
3.
Connect the capacitor to be tested into the multi socket. The
measured value shows on the display.
Note
•
When 1 displays the capacitor is short-circuit or the selected
range is to low.
•
To minimize the measuring error caused by the distributed capaci-
tor, the testing lead should be short as possible.
Transistor test
(See fig 4)
1.
Insert the multi socket into the mA and HzV
Ω
terminals.
2.
Set the rotary switch to the hFE position.
3.
Connect the NPN or PNP type transistor to be tested into the
multi socket. The measured value shows on the display.
Frequency
1.
Insert the red test lead into the HzV
Ω
terminal and the black test
lead into the COM terminal.
2.
Set the rotary switch to an appropriate measurement position in
Hz range.
3.
Connect the test leads across with the object being measured.
The measured value shows on the display.
Replacing the Battery
(see figure 5)
1.
Disconnect the connection between the testing leads and the cir-
cuit under test when battery indicator appears on the display.
2.
Turn the Meter to OFF position.
3.
Remove the screw, and separate the case bottom from the case top.
4.
Replace the battery with a new 9V battery (NEDA 1604 or 6F22
or 006P).
5.
Rejoin the case bottom and case top, and reinstall the screw.
English