43
Performing the test
Fig. 53: Volume measurement with the
Lindab PC 410 and a syringe (up to 100
ml tube volume).
NOTE!
If the volume is smaller or equal 100 ml, a sy-ringe has to be
used. If the volume exceeds 100 ml, a Soot Test Pump will
be needed. The vo-lume of the Soot Test Pump is 163 ml
per strike.
• Close the gas line and seal it with a suitable test plug.
• Switch the Lindab PC 410 on before connect-ing the over-
pressure socket (+) via a hose to the test plug.
• Connect the soot test pump or the syringe with a second
hose and the cross piece.
• Go to the main menu, select "Volume" and confirm with
OK.
• Enter the test volume into the meter according to the
sample volume that has to be drawn with the soot test
pump or the syringe, e.g. 163 ml, that is one strike with the
soot test pump.
WARNING!
Please observe the valid accident prevention regulations!
Fig. 54: Volume measurement with the
Lindab PC 410 and a syringe (up to 100
ml tube volume).