-28-
5.3 CAN Simulation
By the CAN simulation, the data registered in the simulation data table can be transmitted with one touch on
the operation panel.
■ Preparation for Simulation
Click
“Measurement”-> “Measurement setting”
on the menu bar to open the measurement setting screen
and make the following settings.
◆
Selection of the Simulation Function
Select “MANUAL
”
in
“Function”
.
◆
Interface Setting
Set “CAN(High speed)” to the Ch-1 interface in “Interface”.
◆
Registration of the Transmission Data
Select each table (0 to F) in "Simulation data table".
In the settings for each table, select the frame type and set the ID, data, data length, number of repeats,
interval, sweep, etc.
■ Start the Simulation
1. Click “Run measurement
” to open the operation panel.
* The operation panel does not open when measurement is stopped or in the online mode or analog mode.
* Please use the simulation by the trigger because this cannot be used in the communication logger mode.
2. Click the button of the table number to be transmitted.
3. If you want to stop the data transmission for which the repeat is set during transmission, click the “Stop
repeated sending
” toggle button, and then click the table number button for which you want to cancel
transmission in succession. After that, click the “Stop repeated sending
” toggle button to return the but
-
ton to the original state.
■ Stop the Simulation
1. Close the operation panel.
2. Execute “Stop measurement
”. (If you press
Stop measurement without closing the operation panel, the
operation panel will be closed automatically.).
■ Simulation by a Trigger
In the simulation preparation, set the trigger setting to “Send data
” for triggers 0 to 7 in the “
Trigger
”
setting. When measurement is started and the trigger conditions are met, the data transmission control can
be performed automatically. As no operation is required from the operation panel, it can also be used in
communication logger mode.
<
Caution
>
When there is no ACK response from other nodes on the CAN bus or when there is a frame
transmission, the data cannot always be transmitted exactly as the button is clicked.
In addition, the data may not be sent due to factors such as overlapping time of multiple table
settings.
When there is no ACK response, the same data will continue to be sent regardless of the repeat
setting, and in this state the transmission cannot be stopped unless the measurement is stopped.