m) Retain the calibration numbers so that periodic checks for calibration can be made. It is suggested
that a copy of these numbers be retained inside the control enclosure and that a second copy be kept in
files.
n) Reduce the load gradually, and verify that the tonnages displayed by the OmniLink II tonnage monitor
module "track" within one or two percent of those of displayed on the load cells. Failure of this
indicates a non-linearity which could be due to incorrect strain gauge location, improper strain gauge
mounting, or incorrect tie rod tension (in frames of this construction).
o) Press the EXIT softkey and return to the Tonnage Monitor menu.
p) Remove the load cells and associated "stack" elements from the machine.
Calibration is complete.