background image

 

                                                                                                                      Shenzhen Linshang Technology Co., Ltd.

 

4-19-2021        Page  1  of  8   

 

 

 

 

 

 

LS136 

UV Energy Meter 

User Manual V1.0 

 

 

 

 

 

 

 

 

 

 

 

 

Please read this manual carefully before using and reserve it for reference. 

 

 

 

Summary of Contents for LS136

Page 1: ...Shenzhen Linshang Technology Co Ltd 4 19 2021 Page 1 of 8 LS136 UV Energy Meter User Manual V1 0 Please read this manual carefully before using and reserve it for reference ...

Page 2: ...ng the UV intensity and UV energy of ultraviolet light in high pressure mercury lamps halogen lamps and other light sources II Parameters 1 Spectral range 315nm 400nm λp 365nm 2 Irradiance measuring range 0 2000mW cm2 3 Irradiance resolution 0 1mW cm2 4 Energy measuring range 0 999999mJ cm2 5 Energy measuring accuracy 10 5 typical 6 Sampling speed 2048 times second 7 Recording period 32 minutes 8 ...

Page 3: ... environment of 100 for a long time 4 The stored data will not be lost when the power is off and the last test data will be displayed automatically after power on 5 High accuracy the meter can pass the inspection of China Institute of Metrology V Operation 1 Parameter setting In OFF mode long press the POWER button for 3s and enter the parameter setting mode A Trigger mode Auto Manual Short press ...

Page 4: ...an long press the button for quick setting Short press the POWER button to confirm the setting and the meter enters the smooth interface C Smooth 50HZ 60HZ If the UV light source uses AC power supply the frequency of the AC power will affect the accuracy of the power measurement Short press the SELECT button to select the power frequency of the light source select 50HZ for DC power supply 50HZ 50H...

Page 5: ...ment mode in the STOP interface Short press the POWER button and then short press the SELECT button to confirm the new measurement clear the historical data display automatically enter the measurement interface RUN state after 1s The interface dynamically displays energy E real time power value P real time temperature value T measurement time Short press the POWER button or after 32 minutes the me...

Page 6: ...n the following ways 1 Use an USB data cable plug it directly into the USB port of the computer 2 Use the matching 5V switching power supply 3 The mobile power supply for smart phones VI Measurement and Notes 1 The sensor must face the UV light source 2 When not in use please long press the POWER button turn off the meter 3 Avoid contact with corrosive materials and keep away from high humidity 4 ...

Page 7: ... pcs 5 Certificate Warranty card 1 pcs 6 Plastic Case 1 pcs 7 Heat Shield 1 pcs IX Service 1 The meter has one year warranty If the meter works abnormally please send the whole meter to the company for maintenance 2 Provide users with spare parts and lifelong maintenance services 3 Provide the users with the meter inspection service for free 4 Free technical support for long term Manufacturer Shen...

Page 8: ...Shenzhen Linshang Technology Co Ltd 4 19 2021 Page 8 of 8 Website www linshangtech com Service hotline 086 755 86263411 Email sales21 linshangtech com ...