(2) Dimensions 

MC-5100L, MC-5200L, MC-5250L 

 

 
 

MC-5460L, MC-5470L 

 

 
6. Ordering information 

 

M C - 5 1 0 * L   –   M O   –   4 J R 1   A A 0 A 0   –   d d   –   e e   –   f f  

       

           

[1]

   

       

[2]

 

     

[3]

   

   

[4]

 

     

                                         

   

[5]

   

           

[6]

   

           

[7]

 

 
[1] Series model    MC: Mass Flow Controller Series 
    MC-5100L, MC-5200L, MC-5250L, MC-5460L, MC-5470L 

 

*

 

Power connector orientation 0:

 

Side direction, 1:

 

Upper direction 

[2] Valve mode 

   

MO: Normally open (Precision polishing) 

   

MC: Normally close (Precision polishing) 

[3] Fitting size 

   

4VR1:6.35VCR106,    4VR2:6.35VCR124,    4SWL:6.35SWL 

    SPW3: Surface mount 1.1.25W-Seal 79.8mm,

 

SPW4: Surface mount 1.125W-Seal 92mm 

    6JR3: 9.52UJR163.4mm,    4JR4: 6.35UJR156mm (Applied to MC-5460L and MC-5470L) 

   

※ Please consult for more information. 

[4] Optional 

   

B: Applied to liquefied gas,

 

E: Small Range Control (1 to 4SCCM in nitrogen)   

   

 

G:

 

Wide Range specifications (Control range 0.5 to 100 %F.S.) 

    J:

     

Valve is closed at below 2% setting,    M: Applied to ClF3 

   

※ Default setting is labeled “AA0A0”. Please consult for more information. 

[5] Gas type 
[6] Full scale flow rate 
[7] Flow rate unit 

   

SCCM (0 ºC standard), SLM (0 ºC standard) 

 
 

7. Connection 

  (1) Device

 

Net™ connector: Pin type micro connector (IEC60947-5-2)

   

Pin No. 

Signal name 

Function 

Drain 

Shield 

V+ 

Power supply 11 to 25VDC (+) 

V- 

Power supply (-) 

CAN_H 

CAN interface HI 

CAN_L 

CAN interface LO 

 

(2) Rotary switch, LED (on top of the MFC housing)

 

 

 

 
 
 
 
 
 

 

8. Operation 
   

Procedure 

1)

 

This product is packed in a clean room before shipment. Please break the 
seals in a clean room after taking it out of its box. 

 

2)

 

Check the gas type and flow rate, and check the direction of the gas flow 
and the MFC before installation. 

 

3)

 

Check for gas leaks from the tubing with a helium (He) leak detector. 

 

4)

 

Prepare a master for Device

 

Net™ and connect specified power, cable and 

connector. 

5)

 

Connect to the network.

 

6)

 

Turn on power supply and let the equipment warm up for at least 5 minutes 
(Recommended time: 30min).   

7)

 

Please power on the MFC for 30 mins, then conduct zero-point adjustment 

without gas flowing

 by pressing button on the top of device or software.

 

 

8)

 

Device starts controlling flow after inputting setting value by software and 
introducing gas with the specific pressure. 

9)

 

Complete  shut  off  cannot  be  achieved  with  the  mass  flow  controller.  If 
complete shut off is desired, a shut-off valve should be installed.   

10)

 

When  a highly  reactive  gas  is used,  thoroughly purge all  foreign  matter 
from the tubing and the MFC before operation.   

11)

 

When contaminated gas is used, install a filter at the equipment inlet.   

       

 

9. Product warranty 

(1) Period 

This product is guaranteed for a period of 1 year from date of shipment. Defects 
are repaired according to the following regulations. 

(2) Scope 

       

Warranty coverage is restricted to this product only. Any other damage caused 
by this product is not covered. 

(3) Disclaimer facts 
 

       

The following repairs are not covered by the warranty: 

1) Failure caused by by-product of fluid used. 
2) Failure caused by misuse (including careless operation) or incorrect repair or modification. 
3) Failure caused by dropping after purchasing. 
4) Failure caused by a natural disasters. 

 

Even if the warranty period is still in effect, the following items may not be repaired. 

1) When the kind of fluid used in the product is unclear. 
2) The product is returned with fluid remaining inside and safety cannot be 

 

 

confirmed. 

 

The MFC is a precision instrument. Control may become unstable if electric 
noise, temperature change of fluid, pulsation of fluid pressure etc. occurs. 
Please be forewarned. 
This instruction manual is subject to revision without notice. 

 

 
 

                 

http://www.lintec-mfc.co.jp 

Corporate Headquarters 

4-1-23 Sekinotsu, Otsu City, Shiga Pref. 520-2277, Japan

 

TEL.

 

+81-(0)77-536-2210 FAX. +81-(0)77-536-2215 

Tokyo Branch Office

 

3F Hattori Build., 4-30-14 Yotsuya Shinjyuku-ku Tokyo 160-0004, Japan

 

TEL.    +81-(0)3-5366-2801 FAX. +81-(0)3-3341-3513 

NET 

Network Status 

Network Status LED 

MOD 

Module Status 

Module Status LED 

ZERO ADJ. 

Zero Adjustment 

Zero Adjustment button 

Display 

Name 

Description 

Default 

MSD 
LSD 

Node 
Address 

Decimal setting:   
MSD: a digit in tens place 
LSD: a digit in ones place 
The setting number above 64 will be Software Set 
Mode.   

63 

RATE 

Data Rate 

Baud rate switch 
1: 125 kbps 
2: 250 kbps 
5: 500 kbps 
Otherwise: Software set mode 