page >>
17
LBL000097A
LightForce FX User Manual, DCN: O
FX
Hand Piece with Finger Switch
and Small Cone
This treatment head is for general use and applications where fine control
with no contact or soft tissue manipulation during treatment is desired.
This head should be perpendicular to the skin and not approaching at
an angle. The area to be treated and a significant surrounding margin
should be covered evenly in a “painting” or “serpentine” motion. The
Small Cone Head can be used at power settings up to 6 W, 10 cm from
the skin of the patient.
Hand Piece with Finger Switch
and Small Ball
This treatment head is for use in applications where contact and soft tissue
manipulation during treatment are desired. Examples of appropriate
applications for the small ball are muscular or tissue injuries in tight spaces
or near sensitive areas. The small ball is not recommended for use over bone,
open wounds, non-intact skin or areas that would be sensitive to contact.
This head should be perpendicular to the skin and not approaching at
an angle. The area to be treated and a significant surrounding margin
should be covered evenly in a “painting” or “serpentine” motion.
The Small Ball can be used at power settings up to 6 W.
Hand Piece with Finger Switch
and Large Cone
This treatment head is for general use and applications where no contact
or soft tissue manipulation during treatment is desired. This head
should be perpendicular to the skin and not approaching at an
angle. The area to be treated and a significant surrounding margin
should be covered evenly in a “painting” or “serpentine” motion.
The Large Cone Head can be used at power settings up to 15 W,
8 cm from the skin of the patient.
Hand Piece with Finger Switch
and Large Ball
This treatment head is for use in applications where contact and soft tissue
manipulation during treatment are desired. Examples of appropriate
applications for the large ball are muscular or deep soft tissue injuries.
The large ball is not recommended for use over bone, open wounds,
non-intact skin or areas that would be sensitive to contact. This head
should be perpendicular to the skin and not approaching at an
angle. The area to be treated and a significant surrounding margin
should be covered evenly in a “painting” or “serpentine” motion.
The Large Ball can be used at power settings up to 15 W.
Recommendation
Not to be used in a
non-contact mode.
Recommendation
Not to be used in a
non-contact mode.
Hand Piece with Finger Switch
and Flat Window
This treatment head is for general use and applications where fine
control with no contact or soft tissue manipulation during treatment
is desired. This head should be perpendicular to the skin and not
approaching at an angle. The area to be treated and a significant
surrounding margin should be covered evenly in a “painting” or
“serpentine” motion. The Flat Window Head can be used at power
settings up to 6 W, 10 cm from the skin of the patient.