LightForce FX User Manual, DCN: O
FX
Warning
Wear laser safety eyewear provided with the laser before operating the laser device.
Caution
Do not remove the Empower™ Delivery System fiber from the emission port once it has been secured
unless required. Repeated insertion and removal of the fiber will increase the chance of emission port
and fiber tip contamination. If the emission port or the fiber tip is contaminated, the device can be
damaged during laser beam output emission.
The dust cap is installed by the manufacturer as a means to prevent dust and debris from
contaminating the emission port during shipping and before device installation and is not intended
as the primary means to protect the emission port connector during normal use.
We recommend the following procedure:
1. Before using the device for the first time, install and secure the hand piece fiber into the laser
emission port in accordance with section 6.4.
2. Use the device as required.
3. When done using the device, power off and leave the hand piece connected.
4. If the tip is contaminated clean it with the fiber cleaning paper provided with the device.
Bending Limits of the Hand Piece Cable:
The light transferred within the hand piece cable goes through a very small diameter glass rod (an
optical fiber). It can be damaged if bent too sharply. The maximum permissible bending radius of the
optical fiber is 5 cm.
Laser Hand Piece Troubleshooting:
If any part of the hand piece assembly is overheating or producing smoke, then immediately power
off device, discontinue operation, and contact LiteCure® at 302.709.0408 or 877.627.3858 (for US and
Canada) for immediate assistance.
Laser Treatment Head Cleaning:
For information on cleaning refer to section 11.