page >>
26
LBL000097A
LightForce FX User Manual, DCN: O
FX
10
>>
Specifications
11
>>
Cleaning
Model
Laser Type
Laser Wavelength
Aiming Beam Wavelength
Aiming Beam Power
Maximum Output Power
Output port
Operating Modes
Audio warning signal level
Cooling
Dimensions
Weight
Power Requirement
Operation Temperature
Storage Temperature
LightForce® FX
Class 4, Solid State
810
±
20 nm, 980
±
20 nm
650 nm
±
15 nm
3.5 mW
±
1.0 mW
15 W @ aperture of hand piece
Hand Piece
Continuous Wave (CW)
Pulse Wave (50 % duty cycle)
2, 10, 20, 100, 200, 500, 1000, 2500, 5000, 10000 Hz
0 to 65 dB
Thermal Electrically Cooled with Forced Air
14 x 10 x 7 in
|
36 x 25.5 x 18 cm
8.5 lbs
100-240 V ~ 50/60 Hz 250 VA
10
o
C to 32
o
C
-20
o
C to 70
o
C
Caution
DO NOT use Hand Piece Head until alcohol solution used in cleaning procedure
completely evaporates. Doing so may cause the laser to ignite alcohol solution or vapors.
LightForce FX laser device uses solid-state laser technology. It is important to keep the unit and
accessories free from dust. To clean the exterior surfaces on the unit or accessories:
•
Wipe with a soft cloth moistened with isopropyl alcohol solution. Ensure the solution is only 70 %
alcohol or less. Solutions of more than 70 % alcohol can cause product damage.
Warning
Always turn off the system and unplug the power cord from the wall outlet before cleaning.
Warning
Always use protective eyewear and gloves when cleaning and disinfecting any equipment.