Form: CA 045
May 5, 2005
C:\Documents and Settings\timmartin\Local Settings\Temporary Internet Files\OLK9\CA
045 LMI 241-BWV2 to 440 or 501.doc
Page 10 of
12
6
“Master” will be highlighted in screen
header and “G2” is identified as “Gage
Port”. If G2 is not the Gage Port press the
►►
or
◄◄
until G2 appears
7.
Extend the flush tip and retract the gap finger of the LMI 241BWV2. Verify
"Master Low"
highlighted on the collector, press <enter>.
8.
Fully compress the flush tip and extend the gap finger of the LMI 241BWV2. With "Master Hi"
highlighted on the collector press <enter>.
9.
Retract the gap peg so the gage will insert into the “MASTER” gap slot. Gently pull back the base of
the gage so the outside fingers are square in the gap. With "Master Zero" highlighted on the collector
press
<enter>.
Mastering for the 241-BW series is complete.
Gage Port “G2”
“Master” highlighted