CHAPTER 5: MAINTENANCE
5-7
SERIES WS 2 - F190054000
Note
The glass microfibre cartridge cannot be cleaned; replace the entire cartridge when
necessary.
4. EXTRAORDINARY MAINTENANCE
4.1. Centrifugal stage maintenance
Normally the centrifugal stage does not require maintenance, as the design of the impeller makes that
stage self-cleaning with continuous extraction of re-condensed liquids. For some particular applications, it
may be necessary to periodically clean the drainage ring and relative discharge slots following this
procedure:
Release the latches and lift the movable part
unscrew the knob and remove the knob
unscrew the nuts on the impeller case
remove the impeller for cleaning (see Impeller cleaning) only if necessary
clean and remove any deposits from inside the impeller case and the drainage ring
carefully clean the stop flange of the impeller case on the drainage ring
reassemble the impeller case and drainage ring and tighten its nuts
a) Opening
b) Remove fastening nuts
c) Worm screw removal
d) Ring removal
e) Ring cleaning
f) Reassembly
Summary of Contents for FOX WS 2 1020
Page 2: ......
Page 5: ...CHAPTER 1 GENERAL INFORMATION 1 1 SERIES WS 2 F190054000 CHAPTER 1 GENERAL INFORMATION...
Page 12: ...1 8 CHAPTER 1 GENERAL INFORMATION SERIES WS 2 F190054000...
Page 18: ...2 6 CHAPTER 2 TECHNICAL DATA TRANSPORT SERIES WS 2 F190054000...
Page 19: ...CHAPTER 3 GENERAL DESCRIPTION 3 1 SERIES WS 2 F190054000 CHAPTER 3 GENERAL DESCRIPTION...
Page 22: ...3 4 CHAPTER 3 GENERAL DESCRIPTION SERIES WS 2 F190054000...
Page 23: ...CHAPTER 4 INSTALLATION AND USE 4 1 SERIES WS 2 F190054000 CHAPTER 4 INSTALLATION AND USE...
Page 39: ...CHAPTER 5 MAINTENANCE 5 1 SERIES WS 2 F190054000 CHAPTER 5 MAINTENANCE...
Page 50: ...5 12 CHAPTER 5 MAINTENANCE SERIES WS 2 F190054000...
Page 51: ...CHAPTER 6 TROUBLESHOOTING 6 1 SERIES WS 2 F190054000 CHAPTER 6 TROUBLESHOOTING...
Page 55: ...CHAPTER 7 SPARE PARTS 7 1 SERIES WS 2 CHAPTER 7 SPARE PARTS...