During Recording
UTREL30-WiFi User Guide
51
During Recording
During normal operation, the display shows the most recently recorded temperature.
This temperature is updated at the same rate as the logging interval. The current time is
also displayed (in 24 hour format). A tick symbol is shown as long as no alarm event
has occurred. If an alarm event is registered, a cross symbol is shown instead of the
tick.
Below the main temperature the display shows the minimum and maximum
temperatures recorded since the beginning of the trip, or since they were last manually
reset.
At the bottom of the display you can see an alarm day summary, where any days on
which an alarm was recorded are highlighted. The WiFi and cloud connection symbols
are also shown.
Following are some sample display screens:
At 1:25 pm:
l
Alarm events were recorded 6, 18, 19 & 27 days
ago.
The following symbols show on the display:
o
for the alarm that occurred 6 days ago
o
for the alarms that occurred 18 and 19
days ago
o
for the alarm that occurred 27 days ago
l
These alarms were cleared by an inspector, as
the display currently shows the OK tick ( ).
l
The current temperature is over the upper alarm
threshold, as indicated by the upper alarm
marker ( ).
l
The duration of this temperature excursion has
not yet triggered an alarm.