Interpreting the Data
UTREL30-WiFi User Guide
71
Temperature Scale
A temperature scale is shown, which adjusts dynamically depending on the settings
made during configuration of the logger with LogTag
®
Analyzer.
Markers
On the chart, special symbols will mark the readings at which certain events took place:
l
A
º
symbol will be shown, if the UTREL30-WiFi was downloaded with LogTag
®
Analyzer.
l
A
•
symbol will be shown if an inspection mark was placed with the
REVIEW
/MARK
button.
l
A
•
symbol will be shown when the Paused feature was enabled, and any button
was pressed during recording.
l
A
•
symbol will be shown where a reading could not be taken due to USB
communication with a PC.
l
A
•
symbol will be shown when the min/max readings were reset.
USB Paused Readings
This logger cannot take a reading while exchanging data with a PC while being plugged
into a USB port. As a result, you will occasionally see a gap in the graph where the
UTREL30-WiFi would otherwise have taken a reading. The list shows
---
instead of the
reading.
Recorded Data Chart
The chart shows a graphical representation of the data during the trip. As part of the
UTREL30-WiFi configuration process you set the parameters that influence how the
chart is presented.
Legend
Shows the symbols for
download marks
,
inspections marks
and
paused marks
if they
appear in the readings.
File Information
This section shows general information about the PDF file, such as generation time, date
and time formats used in the chart and the data list as well as the file name, which is
compiled from information about the data it contains:
LogTag_[serial_number]_[trip number]_[file creation date]_file creation time]_[OK or ALM].pdf
Other files that may be generated have the extensions *.csv and *.ltd.
Logger Information and Configuration
This section shows general information such as serial number, model number, trip
number, battery status and description. It shows how the logger was started, if a start