33
Guidance and manufacturer’s Declaration
–
electromagnetic
emission
(
SW-1000Sscooter
)
is intended for use in the electromagnetic
environment specified below. The customer or the user of the
(
SW-1000Sscooter
)
should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Complia
nce
level
electromagnetic
environment-guidanc
e
Electrostatic
discharge
(
ESD
)
IEC 61000S-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/ burst
IEC 61000S-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for
power
supply
lines
±1kV for
input/
output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000S-4-5
±1 kV
differential
mode
±2 kV common
mode
±1 kV
differenti
al mode
±2 kV
common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
OWNER’S MANUAL
34
Compact Scooter
Guidance and manufacturer’s declaration –
electromagnetic emission
Limosa scooter is intended for use in the electromagnetic environment
specified below. The customer or the user of the Limosa scooter should
assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compli-
ance level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000S-4-2
Electrical fast
transient/ burst
IEC 61000S-4-4
Surge
IEC 61000S-4-5
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for
power
supply
lines
±1kV for
input/
output
lines
±1 kV
differential
mode
±2 kV common
mode
±1 kV
differenti
al mode
±2 kV
common
mode
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
ver 2019.05.30E