Logs and Reports
209
Problem:
Link corrupted - data corruption has occurred between the detector and LomaLink
PC or other controlling device. The operation of the detector is unaffected.
o
Actions:
1. Check cabling and link connectors from remote device.
2. Check LomaLink PC or other controlling device for reported errors.
Problem:
Link failure - a break in communication between the detector and LomaLink
PC/other controlling device. The operation of the detector is unaffected.
o
Actions:
1. Check cabling and link connectors from remote device.
2. Check connections between the detector and LomaLink PC or other controlling
device for loose or broken wiring/connectors.
Problem: PV test timeout - an automated prompt for the PV test has either not been actioned
or completed within the window time set. The transport system may have stopped.
o
Action:
Clear system fault (restart the transport if stopped) and perform a
PV test
.
Problem:
PEC obstructed - the product registration photo-eye has remained blocked for
greater than 6 consecutive pack lengths. This fault is only generated if the PEC fault is
enabled.
o
Actions:
1. Check the PEC for blockage.
2. Check the PEC is working correctly.
3. Check pack flow length entry is adequate, ensuring an allowance is made for pack
skew.
4. Ensure the correct belt speed is entered.
Problem:
Wrong flash c/sum - the software has compared the flash checksum currently
stored in its memory with the checksum for recently installed software. This occurs if a
software update has been done and a hardware cold start was not performed after
installation.
o
Actions:
1. Perform an update checksum from the tick button found in the
Search Head Setup
screen.
2. Perform a hardware cold start by moving link JP6 across pins 1-2 from 2-3 on the
Control Board, then power on/off This will wipe all system and product setup data from
the detector. Therefore a complete setup of the detector will be required. This should
only be done in the event of extreme difficulties. Return the link to its original position
before entering any data.
Problem:
Drive too low - (IQ3 only) the transmitter drive current is not set.
o
Action:
From the
VF Drive
screen perform the auto configure operation. This will
automatically reset all the drives for each of the detector operating frequencies. This will
take several minutes to complete. This action will not affect any other detector settings.
Summary of Contents for IQ3+ST
Page 2: ......
Page 36: ...Safety First 26 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 43: ...Installation 33 ...
Page 60: ...Installation 50 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 70: ...About the Metal Detector Range 60 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 172: ......
Page 212: ...Operating Instructions 202 THIS IS PAGE INTENTIONALLY LEFT BLANK ...
Page 240: ...Maintenance 230 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 244: ...Glossary 234 THIS PAGE IS INTENTIONALLY LEFT BLANK ...