48
|
Instructions for use
| Reflex Shower-toilet chair
9. Electromagnetic compatibility
Guidance and Manufacturer’s declaration-electromagnetic emissions
The Reflex is intended for use in the electromagnetic environment specified below. The customer or the user of Reflex should
assure that it is used in the following environment:
Emissions Test
Compliance
Electromagnetic Environment - guidance
RF emissions
CISPR11
Group 1
The Reflex uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR11
Class B
The Reflex is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s declaration-electromagnetic immunity
The Reflex is intended for use in the electromagnetic environment specified below. The customer or the user of the Reflex
should assure that it is used in the following environment:
Immunity Test
IEC 60601-1-2 Test
Level
Compliance Level
Electromagnetic Environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be made of wood, concrete, or ceramic
tile. If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s) ±1 kV line(s) to line(s) Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions, and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% Un
(>95% drop in Un)
for 0,5 cycle
40% Un
(60% drop in Un) for
5 cycles
70% Un
(30% drop in Un) for
25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
<5% Un
(>95% drop in Un)
for 0.5 cycle
40% Un
(60% drop in Un) for
5 cycles
70% Un
(30% drop in Un) for
25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. Reflex is
equipped with a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE
Un is the a.c. mains voltage prior to application of the test level