MAT-1 Battery Replacement Tool Kit, Rev A 

 

 

MAT-1 Battery Replacement Tool Kit 

The 

MAT-1 Battery Replacement Tool Kit

 

contains the items that are required to 

replace the battery in a MAT-1 Data 

Logger. The kit also includes one 

replacement battery to get you started. 

Additional batteries may be purchased 

separately. 

Package Contents 

 

Circuit Board Holding Tool 

 

3/32” Hex Driver 

 

Replacement “A” size lithium battery 

Caution

: This logger is sensitive to static electricity when removed from the logger housing.  Hold the 

circuit board by the edges and avoid working on indoor carpet or in other static prone environments. 

Instructions 

1.

 

Remove the end cap of the logger, connect a USB cable and use MAT Logger Commander to 

stop the logger.  Failure to stop a running logger will result in data loss. 

2.

 

Slide the Circuit Board Holding Tool into the end of the logger. Aligning the slot in the tool with 

the circuit board. 

 

3.

 

Locate the two hex screws inside the logger and insert the hex driver into one of them. 

4.

 

Loosen, but do not remove the screw.  Repeat on the other side. 

5.

 

The board should now be loose within the housing. Gently tap the housing and slide the circuit 

board and battery out of the logger.  

 

 

Align 

Circuit Board Holding 

Tool

 and insert fully into 

logger housing.

 

Carefully use the 

Hex Driver 

to 

loosen set screws located 

inside the logger (not shown).

 

Battery 

Desiccants (2) 

USB Port 

Circuit Board Holder 

Set Screws (2)

 

Battery Connector

 

Reviews: