7 | Operation
98
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
After activation of the “Night screen on” button, the night view will come on
within 30 seconds. The display switches back from the night screen to the night
mode if one of the following conditions occurs:
Tapping the screen
30 seconds after the last input the display
switches back from the night mode to the
night screen.
Active alarms
If, after acknowledging an alarm, no alarm oc-
curs for 30 seconds, the display switches back
from the night mode to the night screen.
Manoeuvre in progress
(The night screen remains active dur-
ing automatic manoeuvres such as
“P.01” or “Sigh”.)
30 seconds after the end of the manoeuvre
the display switches back from the night
mode to the night screen.
The night view can only be activated during a ventilation procedure.
If no ventilation is in progress, the Night screen button is dimmed and cannot be
selected.
7.4.5.3
Alarm volume / brightness
Alarm volume
During normal operation (no alarms), elisa 300 / elisa 500 continuously measures
the ambient noise level. When the ambient noise is very loud and an alarm occurs,
the volume of the alarm tone is automatically adapted to the ambient noise level
(Auto Volume ON). The maximum value is 20% above the set value in this case,
but not more than 100%.
However, the volume of the alarm tone will never drop below the preset value.
(A default value for the alarm volume can be preset at the configuration level. If
no alarm volume is preset at the configuration level, the default value will be set
at 80%.)
When you adjust the alarm tone volume, the ventilator will emit a tone with the
actual volume so that a volume suitable for the environment can be selected.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...