9 | Alarms and troubleshooting
110
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
9.1
Problems – causes – corrective measures
Error:
Possible causes:
Corrective measure:
Required reading, ventila-
tion mode, function (e.g.,
Peso), parameter or curve
is not displayed.
The desired function or display
has not been configured for the
user on the configuration level.
Please contact the re-
sponsible person in your
ward or your service
technician.
One of the connected
accessories (CO
2
sensor,
SpO
2
sensor, elisa@megs)
does not work.
The desired function has not
been configured for the user
on the configuration level or it
has been configured for another
interface.
You can review the cur-
rent interface configura
-
tion on the “Status” tab
by selecting the second
view.
If in doubt, please contact
the responsible person
in your ward or your
service technician.
Measured PEEP value is
different from set value.
Valve bar or breathing circuit
were exchanged but no system
test was carried out
Carry out a system test.
Measured PEEP value is
different from set value.
The PEEP membrane is contam-
inated with medication residue af-
ter a drug nebulization procedure
(the expiratory filter is missing
or unsuitable, or the selected
replacement interval is too long
considering the nebulization
intensity or osmolarity of the
applied medication.)
Exchange expiratory
valve
Install suitable expiratory
filter for drug nebuliza
-
tion or choose a shorter
replacement interval.
Ventilation pressure im-
plausibly high
Accessories or combination with
accessories increases the flow
resistance
Exchange breathing
circuit, exchange filter, re
-
move / prevent conden-
sate in breathing circuit
USB stick is not detected
Incorrect or implausible format-
ting.
The elisa 300 / elisa 500
devices support only a
FAT32-formatted USB
stick. The USB stick
needs to be formatted
with a standard operating
system (e.g., Windows).
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...