9 | Alarms and troubleshooting
136
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#248
7
Insp. flow
error
Inspiratory
flow sensor
error
Information:
- The air flow sensor system is defective
- Ventilation is now only possible with oxygen
Decide: If pure oxygen concentration is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
Contact an authorised service agent.
Decide: If emergency ventilation mode is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
- Contact an authorised service agent.
#249
3
Exp. flow
error
Expiratory
flow sensor
error
Information:
- The expiratory flow is only measured up to 80 l/min
- The tidal and minute volume displayed may be
incorrect
Ventilator is operating safely.
7
Information:
- The expiratory flow sensor system is defective
- Tidal and minute volume cannot be measured
- Ventilator now operates in emergency ventilation
mode.
#253
3
Exp. flow
error
Expiratory
flow sensor
error
Information:
- A redundant sensor is defective
- The accuracy of low readings can be affected
Ventilator is operating safely.
#253
7
Exp. flow
error
Expiratory
flow sensor
error
Information:
- The expiratory flow sensor system is defective
- Tidal and minute volume cannot be measured
- Ventilator operates in emergency ventilation mode
Decide: If emergency ventilation mode is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
- Contact an authorised service agent.
#259
3
O2 flow
error
O2 flow
sensor error
Information:
- Two redundant sensors measure implausible values
- The accuracy of the measured values may be
impaired
Ventilator is operating safely.
- Contact an authorised service agent.
#260
3
Air flow
error
Air flow
sensor error
Information:
- Two redundant sensors measure implausible values
- The accuracy of the measured values may be
impaired
Ventilator is operating safely.
- Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...