9 | Alarms and troubleshooting
140
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#315
3
5
Power
failure
The internal
power supply
of the control
unit is defec-
tive.
Information:
- This error does not affect the ventilation function
- Ventilation will continue during change of control
unit
Do the following:
- Exchange the control unit
The new control unit will initialize automatically.
If the alarm recurs or persists:
- Contact an authorised service agent.
#316
3
5
Power
failure
The internal
power supply
of the control
unit is defec-
tive.
Information:
- This error does not affect the ventilation function
- Ventilation will continue during change of control
unit
Do the following:
- Exchange the control unit
The new control unit will initialize automatically.
If the alarm recurs or persists:
- Contact an authorised service agent.
#317
3
5
Power
failure
The internal
power supply
of the control
unit is defec-
tive.
Information:
- This error does not affect the ventilation function
- Ventilation will continue during change of control
unit
Do the following:
- Exchange the control unit
The new control unit will initialize automatically.
If the alarm recurs or persists:
- Contact an authorised service agent.
(*) elisa 300 elisa 500 only (**) elisa 600 elisa 800 only
Power Management Alarms
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#356
3
SpO2
sensor
defective
The SpO
2
sensor reports
an error or is
missing.
Information:
- Sensor cannot be activated
#357
3
SpO
2
signal
quality low
Poor SpO
2
sensor signal
quality.
Information:
- Inadequate SpO
2
signal quality. The measured
values may be implausible
Please check:
- That the sensor is correctly applied to the
finger
- That the skin is not too strongly pigmented
If the alarm recurs or persists:
Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...