9 | Alarms and troubleshooting
142
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#408
1
2
3
4
6
7
Battery >I<
discharged
The battery
in compart-
ment >I< is
discharged.
Delay:
15 seconds max.
Information:
- Battery power is on reserve capacity
- Remaining ventilation time is less than 15
minutes
Do one of the following:
- Restore mains power supply
- Install an additional battery in battery compart-
ment >II<
#409
1
2
3
4
6
7
Battery
>II< dis-
charged
The battery
in compart-
ment >II< is
discharged.
Delay:
15 seconds max.
Information:
- Remaining operating time on battery power is
less than 15 minutes
Do one of the following:
- Restore mains power supply
- Install an additional battery in battery compart-
ment >I<
#410
5
Fan failure
The fan in the
ventilator unit
failed
Do the following:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
If the alarm recurs or persists:
- Contact an authorised service agent.
#411
3
Internal
fault
Communication
error between
ventilator unit
and power sup-
ply module. The
device cannot be
switched off.
Delay:
15 seconds max.
Information:
- Device cannot be switched off
To switch off ventilator:
- Disconnect the mains plug and
- Remove the batteries
If the alarm recurs or persists:
- Contact an authorised service agent.
#412
and
#413
1
Battery
electronics
The battery>I<
(>II<) cannot be
charged.
Delay:
15 seconds max.
Check that:
- Battery is connected correctly
- Battery temperature is not too high
Do the following:
- Reconnecting battery >I< (>II<)
- Wait a few minutes until the battery has
reinitialized itself and the charging process starts
again automatically.
If the alarm recurs or persists:
- Replace the battery in compartment >I< (>II<)
#414
and
#415
1
Battery
electronics
The charging
circuit of battery
>I< (>II<) is de
-
fective. Charging
is not possible.
Delay:
15 seconds max.
Check that:
- Battery is connected correctly
If the alarm recurs or persists:
- Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...