9 | Alarms and troubleshooting
144
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Intensive Care Ventilator Alarms
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#501
4
Expiratory
flow sensor
The heater for
the expiratory
flow sensor is
defective.
Delay:
30 seconds max.
- Information:
- Water condensation may be inside the expirato
-
ry flow sensor
- Expiratory measuring values may be incorrect
Ventilator is operating safely.
If the alarm recurs or persists:
- Contact an authorised service agent.
#502
5
Device tem-
perature
Temperature of
expiratory flow
sensor is outside
the operating
range.
Delay:
10 seconds max.
Information:
- Expiratory measuring values may be incorrect
Ventilator is operating safely.
Check that:
- Ambient temperature is within range
- There are no external sources which are causing
a temperature change
- After transport: let the ventilator adapt to envi-
ronmental conditions
If the alarm recurs or persists:
- Contact an authorised service agent.
#503
5
Device tem-
perature
Temperature
inside the gas
mixer is outside
the operating
range.
Delay:
10 seconds max.
Ventilator is operating safely.
Check that:
- Ambient temperature is within range
- There are no external sources which are causing
a temperature change
- Temperature of supplied gases are within oper-
ating range of ventilator
- After transport: let the ventilator adapt to envi-
ronmental conditions
If the alarm recurs or persists:
- Contact an authorised service agent.
#504
4
ON button
error
The ON button
is stuck or was
pressed too long
Delay:
5 seconds max.
Information:
- Do not press the ON switch longer than 10
seconds
Ventilator is operating safely.
If the alarm recurs or persists:
- Contact an authorised service agent.
#505
(**)
3
Visual alarm
The alarm
indicators of the
ventilator unit
failed.
Information:
- This error does not affect the ventilation
function
- The alarm indicators of the ventilator unit
cannot generate the visual alarms
- Now the control unit generates the visual
alarms
Ventilator is operating safely.
If the alarm recurs or persists:
- Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...