Code Prior-
ity
Alarm
message
Cause
Corrective measure
#506
7
Buzzer fault
The buzzer in
the ventilator
unit is defective
Information:
- This error does not affect the ventilation
function
- The loudspeaker cannot generate the acoustic
alarms
- Now the loudspeaker in the control unit gener
-
ates the acoustic alarm
Ventilator is operating safely.
If the alarm recurs or persists:
- Contact an authorised service agent.
#507
3
Oxygen
sensor
The O
2
meas-
urement in the
gas mixer is
disturbed.
Delay:
5 seconds max.
Information:
- FiO2 sensor calibration in the service menu may
correct the error
If the alarm recurs or persists:
- Contact an authorised service agent.
#508
7
Internal
device
failure
A sensor
communication
failure occurred.
Information:
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
#509
2
Memory
error
An error
occurred while
reading the
default configu
-
ration
Do the following:
- Check all ventilator settings and alarm limits and
adapt them if required
Ventilator is operating safely.
If the alarm recurs or persists:
- Contact an authorised service agent.
#510
7
Internal
device
failure
An internal
voltage supply
has failed
Information:
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
#511
3
CO
2
sensor
defect
The CO2 sensor
reports an error.
Delay:
30 seconds max.
Information:
- Sensor cannot be activated
Do the following:
- Disconnect CO2 sensor from serial interface,
then reconnect
- Replace defective CO2 sensor if applicable
- Use external CO2 monitoring if required
If the alarm recurs or persists:
- Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...