9 | Alarms and troubleshooting
154
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#661
7
Internal fault
The regulation
of the Air flow
control valve is
disturbed.
Information:
- High flow volumes cannot be reliably
provided
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
#663
7
Internal fault
The regulation
of the O2 flow
control valve is
disturbed.
Information:
- Low flow volumes cannot be reliably
provided
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
#664
7
Internal fault
The regulation
of the Air flow
control valve is
disturbed.
Information:
- Low flow volumes cannot be reliably
provided
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
#666
7
Internal fault
A watchdog
security error
occurred.
Information:
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
#668
1
Simulation
mode
No ventilation is
running.
Information:
- To terminate the simulation mode, switch
the device off and switch it on again
#669
(**)
7
Control unit
The software
versions of
control unit and
of the ventilator
unit are incom-
patible.
Information:
- Panel software version mismatch.
Do immediately:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...