Alarms and troubleshooting | 9
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
167
Code
Error message
Corrective measure
#1103
CO
2
sensor calibration
required
Repeat system test
- Calibrate CO
2
sensor
#1104
CO
2
sensor reports error and
cannot be activated
Repeat system test
#1111
to
#1114
Fault safety valve
- Repeat system test
- Check correct assembly inspiratory valve, or
replace valve
- If the problem persists, contact an authorised
service agent