10 | Cleaning, disinfection and sterilisation, disposal
172
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Caution
Check that the device has thoroughly dried before using it again.
10.2.2
Valve bar
D
E
F
10.2.2.1
Removing the three parts of the valve bar
Actuate the respective release lever to remove the individual elements of the
valve bar one after another (push release levers
E
and
F
backwards).
D
Release lever for expiratory flow sensor
E
Release lever for expiratory valve
F
Release lever for inspiratory valve
10.2.2.2
Machine cleaning with disinfection
For machine cleaning and disinfection of the reusable modules of the valve bar, an
EN ISO 15883 compliant washer disinfector can be used. It is suggested to use
the cart for anaesthesia and ventilation accessories for loading the item into the
machine.
Warning
Please note that all valve bar parts may be contaminated by bodily
fluids or exhaled gases!
Warning
Perfect functioning of the valve bar can be guaranteed only if each
individual element is intact. The contact areas of the silicon mem-
branes, in particular, must be carefully checked before each use; the
valves must be replaced if required.
Remark
The three-part
PEEP membrane
(expiratory valve) remains as
-
sembled for reprocessing.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...