Default parameters | 13
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
209
13.1.2
Adjustment ranges
O
2
21–100%
Flow trigger
0.1–20 l/min
Pressure trigger
0.1–10 mbar
PEEP
0–50 mbar
P insp
0–60 mbar
Vt
10–2300 ml
PS
0–100 mbar
Pmax
10–100 mbar
Insp. Ramp
0.05–3 s
Exp Ramp mand
OFF; –100; –80; –50; –20 mbar/s
Exp Ramp spont
OFF; –100; –80; –50; –20 mbar/s
PS Ramp
0.05–2 s
Exp. trigger
0–25%
PS Endflow
5–70%
SwitchFlow
1–80%
Rate
Adults 0–100 /min
Children 0–120 /min
Neonates
0–180 /min
13.1.3
Default alarm limits
When selecting the factory settings, all alarm limits will be set to the least sensi-
tive value.
In the CPR and ALPV modes, different alarm limits are automatically activated.
Refer to the separate tables for the exact values.
Patient category
Adults
Children
MVmin
2.5 l
0.5 l
MVmax
12.0 l
5.0 l
Vtmin (pressure modes)
150 ml
50 ml
Vtmax
OFF
OFF
Leakage (IV)
Leakage (NIV)
50%
95%
50%
95%
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...