Default parameters | 13
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
211
13.1.3.1
Default alarm limits of the ALPV mode
Patient category
Adults
MVmin
= 0.5 * MVi
= 0.5 * (IBW * %MV) / 1000
MVmax
= 2.0 * MVi
= 2.0 * (IBW * %MV) / 1000
Vtmin
= 0.75 * VTi
= 0.75 * IBW * 6.0 ml / kgBW
= 0.75 * IBW * VGF
Vtmax
= 1.60 * VTi
= 1.60 * IBW * 6.0 ml / kgBW
= 1.60 * IBW * VGF
RRmin
= 0.7 * rate
= 0.7 * (%MV / 6.0)
= 0.7 * (%MV / VGF)
RRmax
= 2.0 * rate
= 2.0 * (%MV / 6.0)
= 2.0 * (%MV / VGF)
13.1.3.2
Default alarm limits of the WOBOV mode
Patient category
Adults
Plimit WOBOV
Pmax + 10 mbar