A 2 | Manoeuvres
16
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
In order to prevent volutraumas and barotraumas, the set alarm limits remain
active during delivery of breaths with activated sigh function.
A 2.5
Recruitment manoeuvre
You reach the manoeuvre by pressing the “Manoeuvres” button and then by se-
lecting the “Recruitment” tab.
This function provides a manoeuvre that allows non-ventilated portions of the
lung to be re-opened by elevated ventilation pressures and to keep them open.
The level of invasiveness of the ventilation should be reduced afterwards.
Touch the “Recruitment” tab to open the associated menu.
The user can define the following settings:
• Recruitment time:
set time for recruitment after reaching the inspiratory
pressure (0 to 45 seconds)
• Pinsp:
effective recruitment pressure added to the PEEP (1 to 50 mbar)
• Recr. PEEP:
positive end-expiratory pressure during the recruitment ma
-
noeuvre (0.5 to 50 mbar)
Remark
By directly entering the delta pressure, the manoeuvre can also be
used to implement an open lung strategy. Only the PEEP needs to
be increased for this purpose.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...